Connecting system components

How Can We Help?
Print

Connecting system components

Note: for notations like ‘F2’ please refer to Overview of system components.

Warning!

  • Strictly follow the instructions below. Proper connection of components and modules must be ensured to avoid the possible electrical hazard and/or damage of system components.
  • This system must be properly earthed while using. Improper grounding may possibly cause electrical shock and/or damage of the components.
  • Earth grounding must be checked and validated by authorised electrician before the first use of the system.

To provide proper operation of the system four types of connecting actions need to be established:

  1. Connecting spincube power
  2. Connecting protecting grounds
  3. Connecting spinsplit IoT box
  4. Connecting to electrical network

Connecting spincube power

  1. Connect Chamber Control Connector (F2) located at spinsplit IoT box together with Chamber Control Connector (F10) located at the spincube power.
  2. Plug any end of collector cable to F7 Collector terminal and the other end to D1 Collector terminal.
  3. Push through the high voltage cable of spincube power module of spincube‘s high voltage cable inlet through port D2. Make sure that the cable lies in the energy chain both between the chamber and the power module as well as in the chamber.

Connecting protecting grounds

  1. Connect ground of spincube syringe pump (F17) to spincube power protecting earth terminal (F11).
  2. Connect ground wire of Network Switch module to spincube power protecting earth terminal (F11).
  3. Connect ground wire of spincube chamber to spincube power protecting terminal.

Connecting spinsplit IoT box

  1. Connect USB terminals of USB hub (F23) with USB connector for spincube syringe pumps modules (F18).
  2. Connect USB hub (F23) with spinsplit IoT box using USB.
  3. Connect the LAN connector of spinsplit IoT box (F23) to the Network Switch using CAT5 UTP cable.
  4. Connect the Network Switch to the Local Network (WAN).
  5. Connect the GigE camera LAN connector (F1) to the Network Switch’s Power over Ethernet (PoE) port using CAT5 UTP cable.

Connecting to electrical network

  1. Connect the AC-DC adapters to spincube syringe pump’s power connector (F19). Plug the wall cord of AC-DC adapters into the adapter and to an earthed wall socket.
  2. Plug the mains cord of the Ethernet Switch to an earthed wall socket.
    Caution: To operate this system it is necessary to use a grounded PoE (Power over Ethernet) switch.
  3. Plug in the mains cord of spincube power module’s mains inlet (F9) to an earthed wall socket.
    Caution: proper grounding of the spincube power module must be checked by an authorized electrician before the first use.
Table of Contents