Troubleshooting

How Can We Help?
Print

Troubleshooting

Description Resolution
The HV ON button is pressed, but I am still unable to activate the high voltage. Makre sure that the chamber door is closed.

Make sure that the chamber controller F11 is connected properly yo both: the chamber and the cubepower.

Disabled voltage drops, after the high voltage is activated. Overcurrent indicator is turned on. Decrease the voltage or rearrange the electrospinning setup, e.g., increase the emitter-collector distance.
Droplet formation is not stable – droplet is fluctuating. Decrease the voltage or increase the flow rate.
Droplet size is constantly increasing and drips. Decrease the flow rate or increase the voltage.